【摘 要】
:
Purpose:If rectal balloon were not used in rectum cancer treatment,there is a lack of reproducibility of rectum shape on each treatment and irregularly inserted air inside the rectum can cause inhomog
【机 构】
:
Department of Bio-Convergence Engineering,Korea University
【出 处】
:
The 15th Asia-Oceania Congress of Medical Physics (AOCMP2015
论文部分内容阅读
Purpose:If rectal balloon were not used in rectum cancer treatment,there is a lack of reproducibility of rectum shape on each treatment and irregularly inserted air inside the rectum can cause inhomogeneity of density with surrounding tissue,which could be resulted in uncertainty of dose delivery.In this research,we estimated the accuracy of TPS(Treatment Planning System)dose calculation and actual dose,and compared dose distribution when homogeneity of rectum and surrounding tissue increases as inner materials of rectal balloon changes.Materials and Methods:Cylindrical PMMA RPLGD phantom was used to measure dose at rectal wall and Air,water and PMMA were used as inner balloon materials.Total 12 plans(prostate and rectum cancer 3D-CRT(3 Dimensional Conformal Radiation Therapy)and IMRT(Intensity Modulated Radiation Therapy)plans with air,water,and PMMA balloons)were made using Varian Eclipse ver.8.9 to estimate influence of rectal balloon material on dose distribution.Each 16 glass dosimeters were located at rectum wall so that the point doses were compared with TPS.Result:As Homogeneity increase,Dose distribution was improved.As CT HU value of inner balloon material increase from 0(air)to 1000(water)and 1120(PMMA),Homogeneity Index(HI)and Conformity Index(CI)were increased about 20%and 5%,respectively.And considerable difference,between TPS and RPLGD reading,was measured with air balloon when water and PMMA balloons had less variation.Conclusion:It was shown that the dose distribution and the accuracy of TPS were better when the density of the balloon material was similar to the density of surrounding tissue.Especially when air is inserted into rectum,there is a possibility of difference between actual dose and TPS calculation.Thus,it is needed to look forward to find a method to increase treatment accuracy using tissue-equivalent inner balloon materials.
其他文献
To estimate the α/β ratio of late rectal toxicity in case of prostate cancer radiotherapy; and to investigate whether correcting the rectal dose-volume histogram(DVH)for differences in dose per fracti
Objective To compare the dosimetric difference between anisotropic analytical algorithm(AAA)and pencil beam convolution(PBC)for intensity modulated radiation therapy(IMRT)Plans in esophageal carcioma.
Purpose:The efficacy of spatially fractionated(GRID)radiotherapy for treating bulky tumors has been recognized and demonstrated.GRID therapy might improve tumor control while reducing normal tissue to
Purpose:Cone-beam x-ray luminescence computed tomography(CB-XLCT)opens new possibilities to perform molecular imaging with x-ray.The spatial resolution of CB-XLCT was limited because of severe ill-pos
Purpose:The purpose of this study was to test feasibility of clinical usage of flattening filter free(FFF)beam for treatment of lung stereotactic ablative radiotherapy(SABR).Methods:Ten patients were
Purpose:End-of-exhale(EOE) phase is generally preferred for gating window because tumor position is more reproducible.However,other gating windows might be more appropriate for dose distribution persp
Purpose:To investigate the cumulative dose delivered to the High-risk clinical target volume(HR-CTV)throughout the course of cervix cancer treatment with external beam radiotherapy(EBRT)and brachyther
Purpose:PET visualization method is routinely used in clinical practice,however,it has high interobserver variations and relies on the experience of nuclear medicine physician.PET automated threshold
Purpose:To quantitatively evaluate the strategic innovation and accuracy variation of deformation registration algorithm for parotid glands using the similarity Dice coefficient during the course of r
Purpose:To improve an in-house fully automatic treatment planning system for radiotherapy of all kinds of cancer.Methods:The in-house autoplan system is a patent expert system designed to be embedded