103型色谱仪的 FID(氧焰离子化检测器)在使用中常见故障之一是,点火线圈点火时,因高温而使发射极的固定脚氧化锈蚀、松动及电极表面形成的氧化薄膜弹起或脱落,结果信号输出不稳。解决措施是用金相砂纸对锈蚀部分打磨,再重新安装。为了避免出现此故障,平时应采取预防措施,点火后立即切断电源;控制气体流量和进样量,减少火焰对电极的烘烤。在结构上,可将点火线圈改为图(b)的结构,而点火线圈的固定脚可如图(a)所示加以延长,电极材料改为不锈钢。
103 FID (oxygen flame ionization detector) in the use of common fault is that when the ignition coil ignition, due to high temperature and the emitter of the fixed foot oxidation corrosion, loose and oxide film formed on the electrode surface bounce Or fall off, resulting in unstable signal output. The solution is to use metallographic sandpaper on the rust part polished, and then re-install. In order to avoid this failure, usually take precautionary measures, cut off the power immediately after ignition; control gas flow and injection volume, reduce the baking of the flame on the electrode. Structurally, the ignition coil can be changed to the structure of Fig. (B), and the ignition coil can be extended as shown in Fig. (A). The electrode material is changed to stainless steel.