故障现象曝光时按下曝光手闸第一档,就会发生连接的配电柜跳闸。故障分析与维修将高压发生器内旋阳极插头拔下,用跳线将温度反馈的两个点短接起来(目的是断开旋阳极负载),按下曝光手闸第一档,未出现漏电保护器跳闸现象,将旋阳插头接好,故障现象再次出现。用万用表测量旋阳级的两个线圈,阻值正常(大约几欧姆),判断是旋阳级转子抱死导致电流过大跳闸。在使用varian(瓦里安)的球管该故障率较高,更换球管,曝光一切正常,设备恢复正常。
Fault phenomenon Exposure exposure hand brake first gear, it will happen connected to the distribution cabinet trip. Failure Analysis and Maintenance Unplug the positive anode plug of the high voltage generator, jumper short the two points of the temperature feedback (the purpose is to disconnect the anode load), press the exposure of the first gear of the hand brake, there is no leakage Protector tripping phenomenon, the positive plug will be connected to the sun, the fault phenomenon again. Use a multimeter to measure the two coils of the gyro-level, the normal value of resistance (about a few ohms), determine the rotor is too yaw-shaped lead to excessive current tripping. In Varian tubes, the failure rate is higher, the bulb is replaced, the exposure is normal, and the device returns to normal.