目的对三种不同型号全自动血细胞分析仪手动与自动模式的结果进行比对,分析结果之间的偏差及相关性,提高血液标本检测的准确性。方法随机选取高、中、低值标本共40例分别在Sysmex XE-5000,Sysmex XN-2000及迈瑞BC-6800全自动血细胞分析仪进行检测,同一仪器手动与自动模式共60例新鲜全血标本进行比对,对各仪器及仪器不同模式的白细胞(WBC)、红细胞(RBC)、血红蛋白(HB)、红细胞压积(HCT)及血细胞板(PLT)结果进行相关性分析。结果检测结果均在可允许误差范围内,相关系数>0.98,P>0.05,Sysmex XE-5000,Sysmex XN-2000及迈瑞BC-6800检测结果无显著性差异。结论 3台不同型号血细胞分析仪检测血常规结果之间存在可比性,可交替用于检验工作。
OBJECTIVE To compare the results of manual and automatic modes of three different types of automatic hematology analyzers, to analyze the deviation and correlation between the results and to improve the accuracy of blood samples. Methods A total of 40 samples with high, medium and low value were randomly selected for testing in Sysmex XE-5000, Sysmex XN-2000 and Mindray BC-6800 automatic hematology analyzer. 60 fresh whole blood samples The WBC, RBC, HB, HCT and PLT results of different instruments and instruments were analyzed. Results The test results were within the allowable error, the correlation coefficient> 0.98, P> 0.05, Sysmex XE-5000, Sysmex XN-2000 and Mindray BC-6800 test results no significant difference. Conclusion There are comparability between the results of blood tests of 3 different types of hematology analyzers, which can be used alternately for testing.