故障现象:在仪器计数过程中,发现血红蛋白项目有AIRBUBBLES报警讯号,此时该仪器不能继续正常工作。故障分析与处理:造成该仪器报警信号原因通常为:①稀释液已用完;②因管路压扁致使血红蛋白液无法吸入检体;③可能电位偏移;④可能比色管道内部脏所致等。首先排除稀释液
Symptom: During the counting of the instrument, it was found that the hemoglobin item had an AIRBUBBLES alarm signal. At this time, the instrument could not continue to work normally. Failure analysis and processing: cause the instrument alarm signal is usually caused by: 1 dilution has been used up; 2 due to line flattening caused by hemoglobin fluid can not breathe into the sample; 3 may be potential shift; 4 may be due to internal colorimetric pipeline dirty Wait. First eliminate the diluent