全身照射时要求照射剂量均匀,如果全身照射剂量分布不均匀,会造成病灶复发或增大对正常组织不应有的损伤.在SATURNE 42直线加速器用6MV X线开展全身照射的剂量分布情况研究,加上补偿器后水中 测量均匀度为±2.3%,剂量率用调节加速器增益方式实现0.045Gy/min的要求.在仿真人体模上,用热释光元件所测量的均匀度在±5%以内(极大点头顶点、肩顶点无散射吸收除外),这样的结果完全能满足临床全身照射(TBI)的剂量要求.
Whole body irradiation requires uniform irradiation dose, if the whole body irradiation dose distribution is not uniform, will cause the lesion recurrence or increase the damage to the normal tissue.In the SATURNE 42 linear accelerator with 6MV X-ray to the whole body irradiation dose distribution study, After adding the compensator, the measurement uniformity in water is ± 2.3% and the dose rate is adjusted to 0.045Gy / min by adjusting the accelerator gain mode. On the simulation manikin, the uniformity measured by the thermoluminescence element is within ± 5% (Extremely nodding apex, except at the apex of the shoulder with no scattering absorption), the result of which is fully satisfactory for the dose required for clinical total body irradiation (TBI).