针对现行环境试验设备温湿度校准规范JJF 1101—2003《环境试验设备温度、湿度校准规范》中存在的不合理之处提出建议,对规程规定的测试点分布提出合理化建议,在温湿度偏差、温湿度均匀度、温湿度波动度等校准项目基础上增加了波动最大测试点的温度波动度、温度过冲量、温度过冲恢复时间、各测试点实际温湿度、升降温速率等校准项目,更加全面合理的表征了环境试验设备的各项技术指标,为用户提供了更全面的技术支持。
In view of the existing unreasonable place in temperature and humidity calibration specification JJF 1101-2003 of “Environmental Test Equipment Temperature and Humidity Calibration Standard”, the author puts forward reasonable suggestions on the distribution of test points stipulated in the regulations. In temperature and humidity deviation, temperature Humidity uniformity, temperature and humidity fluctuations on the basis of the calibration items such as increased temperature fluctuations maximum test point temperature fluctuations, temperature overshoot, temperature overshoot recovery time, the test point of the actual temperature and humidity, heating and cooling rate calibration project more comprehensive Reasonable characterization of the environmental test equipment of the technical indicators, to provide users with more comprehensive technical support.