F30-IIB型200mAX线机,功能较全,操作方便,但该机电子限时器的故障率较高,直接影响X线摄片感光量以及病人和工作人员的健康。当x线发生后,不能及时关闭时,又可直接危及球管、高压变压器等的安全。因此,对该机电子限时器发生的故障加以分析具有现实意义,现将其分述如下。一、曝光时间普遍明显缩短,胶片感光不足检查:卸下高压变压器初级的保险丝,打开控制台后盖,对控制台通电,进行摄片操作。除各档时间明显缩短外,各继电器动作次序正常。用万用表测得C_2(电源滤波电容)二端有250V电源电压.用R×10K档测C_3(限时充电电容)充放电正常,无极间短路现象。测电阻群各电阻阻值,发
F30-IIB 200mA X-ray machine has more complete functions and is easy to operate. However, the machine’s electronic timer has a higher failure rate, which directly affects the sensitivity of the X-ray film and the health of patients and staff. When the x-ray is generated and it cannot be closed in time, it can directly endanger the safety of the tube and high-voltage transformers. Therefore, it is of practical significance to analyze the failure of the electronic time limiter of the machine. First, the exposure time is generally shortened significantly, insufficient film photosensitivity check: remove the fuse of the high voltage transformer primary, open the back cover of the console, power on the console, and take a photo. In addition to the obvious shortening of the time of each file, the order of actions of each relay is normal. Measured with a multimeter C_2 (power filter capacitor) two ends of the 250V supply voltage. With R × 10K file measured C_3 (limited time charging capacitor) charge and discharge normal, no short-circuit between the poles. Measure the resistance of each resistance group, send