采用Al_2O_3:C晶体制备了光释光剂量计(OSLD),并开展了剂量学性能的初步研究,结果表明:剂量计退火40 min后本底为0.03 mGy,对10片剂量计分别进行7次重复测量,测量结果的相对标准偏差为0.8%~5.3%,剂量计在0.02 mGy~10 Gy范围内线性响应良好(图1),照后两个月信号衰退小于5%。OSLD具有元件测读过程相对简单、性能稳定、受环境温湿度影响小等特点,在个人剂量监测、环境辐射监测、空间辐射测量、辐
The photoluminescence dosimeter (OSLD) was prepared by using Al_2O_3: C crystal and the dosimetry performance was studied. The results showed that the background of the dosimeter was 0.03 mGy after annealing for 40 min, Repeated measurements showed a relative standard deviation of 0.8% -5.3%. The dosimeter had a good linear response in the range of 0.02 mGy to 10 Gy (Fig. 1), with a signal decline of less than 5% two months after irradiation. OSLD has the advantages of relatively simple device reading and measuring process, stable performance and small influence on the temperature and humidity of the environment. In the personal dose monitoring, environmental radiation monitoring, space radiation measurement,