YKZ-4B型牙科综合治疗机是一种新型的口腔科综合治疗设备。配有高速车头,低速直弯车头等,车头系统采用气动控制,功能较多,在各级医院中广泛应用。我院两台YKZ-4B型牙科综合治疗机使用一年后,经常发生气路、水路阻塞和漏气、漏水故障。经检修分析,发现这类故障大多数发生在气控气阀和气控开关上,如气控阀内的橡胶板老化失去弹性,油污粘结使其不通畅,气控水阀内的弹簧生锈失去弹性致使阀不通畅等。
YKZ-4B-type dental treatment machine is a new type of dentistry comprehensive treatment equipment. Equipped with high-speed front, low-speed straight bend front, the front system pneumatic control, more functions, widely used in hospitals at all levels. A YKZ-4B dental treatment unit in our hospital after one year of use, often occur gas, waterway obstruction and leakage, leakage fault. After inspection and analysis, found that most of these failures occurred in the gas-controlled valves and gas-controlled switch, such as pneumatic valve rubber sheet aging loss of elasticity, oil adhesion makes it unobstructed, pneumatic valve spring rust Loss of elasticity causes the valve to be unobstructed.