论文部分内容阅读
故障现象:在仪器检测过程中,出现cuv·cannot entermeas·block报警信号,此时仪器不能继续正常工作。 故障分析与处理:造成该仪器报警信号原因通常有以下几种情况:①传输系统中,齿轮与比色杯托架不吻合,使传输发生机械故障;②电磁阀棒失灵;③托架弹出器是否完全弹出托架,等等。 首先,按回车确认“Eject rack”,使传输系统移液位置的所有比色杯托架立即被弹出,在托架传输过程中发生机
Symptoms: In the instrument detection process, the cuv·cannot entermeas·block alarm signal appears, and the instrument cannot continue to work normally. Fault analysis and processing: Causes the alarm signal of the instrument to be caused by the following conditions: 1 In the transmission system, the gear and the cuvette tray do not agree with each other, causing mechanical failure during transmission; 2 Failure of the solenoid valve stem; 3 rack ejector Whether the tray is fully ejected, and so on. First, press Enter to confirm the “Eject rack” so that all cuvette trays in the pipetting position of the transport system are immediately ejected and the machine is generated during carriage transfer.