本文报告了医用诊断X射线透视机受检者入射空气比释动能率的测量结果。在159台被测机中有19台超过了国家规定的50mGy·min-1标准,占11.9%。超标原因与X射线机球管窗口处的固有过滤板厚度不足等因素有关。
This article reports the measurement of kerma rate of incident air from a medical diagnostic X-ray machine. 19 of the 159 tested machines surpassed the national standard of 50mGy · min-1, accounting for 11.9%. Exceeding causes and X-ray tube window at the inherent filter board thickness and other factors.