测量了角响应、径向微分响应、能量响应及球腔内中子计数率分布,并估算了有效中心。角响应小于5%,径向微分响应与实心球探测器的情况相同。在源距大于5倍计数器半径的情况下,忽略有效中心与几何中心之差异所导入的响应值误差小于0.7%。球腔内超热与热中子计数比小于10%,计数率依赖于热中子探测器所在位置。大约在200eV至5MeV能区,能量响应是平伸的,误差小于±18%。
The angular response, radial differential response, energy response and intracavity neutron count rate distribution were measured and the effective center was estimated. The angular response is less than 5% and the radial differential response is the same as for solid ball detectors. In the case of a source distance greater than 5 times the radius of the counter, the discrepancy of the active center with the geometric center is ignored and the response error introduced is less than 0.7%. The ratio of hypersphere to thermal neutron in the sphere is less than 10%. The counting rate depends on the position of the thermal neutron detector. About 200eV to 5MeV energy region, the energy response is flat, with less than ± 18% error.