原厂生产的 YC1900—8型医用高压氧舱虽然设计结构在国家标准允许范围内,但舱内设有强电电器设施,存在潜在的不安全因素,舱内一但发生火灾,将造成不可估量的后果。故对氧舱作如下的性能改进。1 空调系统原设计空调系统,采用分体壁挂式双温空调器,空调室内机设在舱内。除将空调室内机控制部分移至舱外操作台处,舱内空调室内机还余有220V、16W 风扇电机一台,温度感应热敏电阻一个。改
Although the YC1900-8 Medical Hyperbaric Oxygen Chamber produced by the original manufacturer is within the allowable range of national standards, there are strong electrical and electronic facilities in the cabin. There are potential unsafe factors. Once a fire occurs in the cabin, it will cause incalculable damage. s consequence. Therefore, the performance of the oxygen chamber is improved as follows. 1 The original design air-conditioning system of the air-conditioning system adopts a split wall-mounted dual-temperature air conditioner, and the air-conditioning indoor unit is located in the cabin. In addition to moving the air-conditioning indoor unit control part to the outboard operation console, there are 220V and 16W fan motors and one temperature-sensing thermistor in the cabin air-conditioning indoor unit. change