压力蒸气灭菌化学指示带是用于辅助监测灭菌效果的,但我们在实践中发现,尽管在压力蒸气锅内物品的摆放,灭菌的时间、温度、压力均符合标准要求,甚至生物监测也完全符合标准要求时也有指示带变色不全的情况出现。为此作者对1997年5月-7月出现的72件变色不全情况进行了调查,确定影响变色不全的原因是指示带受潮、过期、粘贴位置不当、被晒和打包不规范。对此提出了一些如何正确存放和使用指示带的建议,改进后未再出现变色不全
Pressure steam sterilization chemical indicator tape is used to help monitor the sterilization effect, but in practice we found that despite the placement of items in the pressure steam pot, sterilization time, temperature, pressure are in line with standard requirements, and even biological Monitoring also fully meet the requirements of the standard when there is incomplete discoloration occurs. For this reason, the author investigated 72 cases of discoloration incomplete occurred in May-July 1997, and determined that the reason of incomplete discoloration was that the indicator belts were damp, overdue, improperly placed and were not standardized in the sun and packaged. In this regard, some suggestions on how to properly store and use the indicator belt are proposed.