我院子1983年9月中旬在病房、门诊观察室连续发生多起输液反应,现将调查经过报告如下:3例患者均在输入约200ml 左右液体后出现输液反应,经查所用3批号输液均符合规定,输液操作按常规,但消毒输液胶管用无锡1970年生产的手提式高压蒸汽消毒器温度进行测试时,压力表显示20磅/时~2(126.5℃)消毒20分钟,置消毒器内3种温度指示剂中,除安替比林外,苯甲酸和升华硫
I yard in mid-September 1983 in the ward, outpatient observation room occurred consecutive infusion reactions, the investigation after the report is as follows: 3 patients were about 200ml after infusion of liquid infusion reaction, the investigation by the three batches of infusion are in line with The regulations, infusion operation according to conventional, but sterile infusion hose produced in Wuxi in 1970 portable high pressure steam sterilizer temperature test, the pressure gauge shows 20 pounds / hour ~ 2 (126.5 ℃) disinfection for 20 minutes, set the sterilizer 3 In the temperature indicator, in addition to antipyrine, benzoic acid and sublimation of sulfur