来源 :医疗卫生装备 | 被引量 : 0次 | 上传用户:xuhaoumsl
下载到本地 , 更方便阅读
声明 : 本文档内容版权归属内容提供方 , 如果您对本文有版权争议 , 可与客服联系进行内容授权或下架
论文部分内容阅读
故障现象 机器自检正常,但在测量时出现TROVBLE—4。故障分析与检修 打印TROVBLE—4,提示仪器光强度不足。首先检查钨灯灯丝未断,灯丝电压DC12V正常,更换灯泡,故障依旧,怀疑内部有元件损坏。由于仪器没有电路原理图,给维修带来了困难。仔细分析线路结构,根据故障现象,怀疑故障出在光电转换及前置放大部分。用万用表测量前置放大板的三路输出9、7、5脚,在测量状态下,量钨灯点亮期 Symptoms The machine self-test is normal, but TROVBLE-4 appears during the measurement. Fault Analysis and Inspection Print TROVBLE-4, indicating that the light intensity of the instrument is insufficient. First check the tungsten filament is not broken, the filament voltage DC12V normal, replace the bulb, the fault remains the same, suspected internal component damage. Because the instrument does not have a circuit schematic, maintenance is difficult. Careful analysis of the line structure, according to the phenomenon of failure, suspected of failure in the photoelectric conversion and preamplification. Use a multimeter to measure the three-way output of the preamplifier board 9,7,5 feet, in the measurement state, the amount of tungsten lighting period
其他文献
适应社会主义市场经济体制,适应新的医学模式转变,适应人民群众多层次的医疗卫生保健需求(以下简称“三个适应”),是我国乃至今后一个时期卫生改革与发展的方向,也是具有革命性意义
4.一次性用品的广泛使用。目前医院在治疗和手术中普遍使用一次性用品,它包括卫生、医疗器材、生活用品等,使用的面从治疗扩大到检查,从护理扩大到生活从病人扩大到医务人员,这些都
按照(学校卫生工作条件)的要求,我市每年在9~10月组织各地段医院对所辖区的中小学生进行一次健康检查,由于受检学生多,工作量大,有些检查人员为求进度而忽视质量,使检测误差发生率很
贝克曼700生化分析仪用99程序检修二例山东省滨州地区中心医院陈冬生,王如芹,赵波例一故障现象:稀释器的两注射器不吸液,仪器无法对样本稀释。操作主机:主机显示操作程序,但稀释器不动作
1 故障现象 主监视器显示图象模糊,副监视器不亮。2 故障检修 根据故障现象,首先操作面板上各键钮,各控制键钮均能正常起控。将灵敏度调到最大,超声图象仍然模糊,调节主监
实行中小学教材出版发行招标是我国推进教材出版发行改革,减轻学生家长负担,打破垄断,引进竞争机制的一项重要举措。根据国务院办公厅有关精神以及新闻出版总署、教育部和国
浅谈医院药学资料室的建设邱业功,朱岩华(新疆维吾尔自治区人民医院乌鲁木齐830001)医院药学的现代化使医院药学资料室的建设受到重视。医院药剂科建立资料室的使命是:①对医师使用新药
经新闻出版行政主管部门批准,自今年6月1日起,本刊由原半月刊改为旬刊,即原上半月版改为上旬版,每月1日出刊;原下半月版改为下旬版,每月21日出刊;今天—6月11日出版的本刊,为新创办的
抓住机遇深化改革实现卫生事业发展新目标──“九五”卫生事业发展与改革的思考江西省卫生厅厅长周标“八五”期间,我省各级卫生部门不断深化卫生改革,主动适应社会主义市场经
请下载后查看,本文暂不支持在线获取查看简介。 Please download to view, this article does not support online access to view profile.