A novel technique for Correction of magnetic field inhomogeneity of MRI cardiac T2*

来源 :The 15th Asia-Oceania Congress of Medical Physics (AOCMP2015 | 被引量 : 0次 | 上传用户:shangxiao15
下载到本地 , 更方便阅读
声明 : 本文档内容版权归属内容提供方 , 如果您对本文有版权争议 , 可与客服联系进行内容授权或下架
论文部分内容阅读
  Introduction:Cardiac T2* becomes a standard method to assess iron accumulation in myocardium.The reliability of T2* during the course of a longitudinal study could be affected by the variation of magnetic field inhomogeneity.We propose a novel technique for the cardiac T2* correction when the magnetic inhomogeneity is entangled.Objective:This study aims to find a correction method for cardiacT2* when the main magnetic field inhomogeneity was involved during the course of treatment.Materials and method:A stable reference(SR)T2*phantom filled with 20 different MnCl2concentrations(equivalent to the range of T2* from 1.31 ms.to 44.90 ms.)was scanned prior to 2 volunteers' scan.The second set of reference(SSR)phantom including 3 levels of T2* 9.06,13.63,20.31 ms,were placed to cover anterior body of the sternum and were scanned with short-axis view of the mid left ventricle of the 2 volunteers.Images were acquired with 10 echo times using the multi-echo gradient echo pulse sequences.Four sets of images were acquired including one control and 3 sets of images at 3 different locationsofmagnetic field perturbations.All scans were performed on a 1.5 T Achieva,Philips MRI scanner.The changes of the T2* of the SSR phantom due to the field perturbation were used to correct for the cardiac T2* by a piecewise quadratic interpolation fromthe SR phantom data.Results:Average cardiac T2* of the 2 volunteers as a control were 13.08 ms and 17.80 ms.Average cardiac T2* of the 3 different locations of field perturbation changed to 12.31 ms and 16.99 ms.When applied the novel correction techniques,the T2* of the 2 volunteers were 13.04 ms.and 17.77 ms.Conclusion:This new technique promisingly reduced the effect of magnetic inhomogeneity for the cardiac T2* measurements and provided more accurate T2*.
其他文献
  Purpose: To propose a method for adaptively optimizing an image quality with over luminance in order to realize the optimization of visual quality of over l
会议
  Purpose: This paper reports the relationship between Body Mass Index(BMI),tube voltages,tissue weighting factors,gender and the conversion factors of DLP,th
会议
  Background and purpose: The effect of methamphetamine(MA)abuser on human brain morphology has been reported.Several studies show evidence of structural brai
会议
  Purpose:Low count PET studies suffer from quantitative bias in iterative reconstruction methods such as MLEM or OSEM.This work aims at reducing PET image bi
会议
  Purpose Phase retrieval methods are used to disentangle pure phase from the phase contrast images.Thereare many phase retrieval methods available that addre
会议
  Purpose To investigate the dosimetric characteristics of TLD2000 thermoluminescent dosimeter(TLD).Methods 350 TLD2000 TLDs were screened to have the same se
会议
会议
  Aims:To compare flattening filter free(FFF)beams of tomotherapy(TOMO)and C-arm based flattening filter free beams in volumetric modulated arc therapy(VMAT)o
会议
  Time consuming of dynamic multileaf collimator(dMLC)quality assurance(QA)is an issue for clinical radiotherapy department.Considering the time consuming pro
会议
  Purpose:Respiratory gated radiation therapy(RGRT)gives accurate results when a patient's breathing is stable and regular.Thus,the patient should be fully aw
会议