【摘 要】
:
Purpose To investigate the dosimetric characteristics of TLD2000 thermoluminescent dosimeter(TLD).Methods 350 TLD2000 TLDs were screened to have the same se
【机 构】
:
Peking University Third Hospital
【出 处】
:
The 15th Asia-Oceania Congress of Medical Physics (AOCMP2015
论文部分内容阅读
Purpose To investigate the dosimetric characteristics of TLD2000 thermoluminescent dosimeter(TLD).Methods 350 TLD2000 TLDs were screened to have the same sensitivity within ±3.0%.TLDs were irradiated to evaluate the response repeatability and dose rate effect using 6 MV X-rays and I-125 seed.TLDs were irradiated to different doses using 137Cs(662 keV γ-rays),125I seed and 6 MV X-rays,to evaluate the dose response of the TLDs.TLDs were irradiated to the same dose using 125I seed,137Cs,48,65,83,118,250 keV and 6 MV X-rays,to study the energy response of the TLDs.Results The maximum deviations of the repeatability were 2.7%and 4.0%for 6 MV X-rays and 125I seed,respectively.No dose rate effects were observed for TLD2000 TLDs.The dose response of TLDs for 137Cs and 125I seed were linear.For 6 MV X-rays,the linear response range was 0.74-10.0 Gy.The dose response became supralinear when the irradiated dose was beyond 10.0 Gy.The energy response for 125I seed,48,65,83,118,250 keV and 6 MV X-rays,relative to the energy response of 137Cs were 1.70,1.25,1.08,0.99,0.91,0.96 and 1.22,respectively.Conclusions TLD2000 TLD has good repeatability and linear dose response for 137Cs,125I seed and 6 MV X-rays without dose rate effect,but the dose response is energy dependent.
其他文献
Objective:Evaluated the speed and accuracy of GPU-based collapsed cone dose calculation.Methods:Ten Head and Neck patients and Ten cervical carcinoma patien
Objective:To investigate the treatment plan quality between two treatment planning systems(TPS)for volumetric modulated Arc therapy(VMAT).Materials and Meth
Purpose: The application of limited-angle computed tomography(CT)has great expectation in some clinical circumstance such as dental CT imaging.Existing iter
Purpose:Head motion in PET/CT leads to artifacts due to the mismatch of PET and CT images.This work aims at reducing motion artifacts in PET image reconstru
Purpose: To propose a method for adaptively optimizing an image quality with over luminance in order to realize the optimization of visual quality of over l
Purpose: This paper reports the relationship between Body Mass Index(BMI),tube voltages,tissue weighting factors,gender and the conversion factors of DLP,th
Background and purpose: The effect of methamphetamine(MA)abuser on human brain morphology has been reported.Several studies show evidence of structural brai
Purpose:Low count PET studies suffer from quantitative bias in iterative reconstruction methods such as MLEM or OSEM.This work aims at reducing PET image bi
Purpose Phase retrieval methods are used to disentangle pure phase from the phase contrast images.Thereare many phase retrieval methods available that addre