论文部分内容阅读
在工作中比较了盛装注射器用铝饭盒与自动启闭式消毒盒(空军后勤部门诊部研制)的灭菌效果。测试时,在盒中放化学指示卡(山东新华医疗器械厂)以反映灭菌效果。测试的容器,分上、中、下3层,每层在4角与中央按梅花状放5个点,每次试验共15个点。灭菌温度为121℃,作用20分钟。铝饭盒测试32次,共30盒次,30张指示卡有12张由浅黄变黑,合格率为40%。自动启闭式消毒盒测试了10次,共150盒次,
At work, compared with the aluminum containers filled with syringes and automatic opening and closing disinfection box (Air Force logistics department clinic developed) sterilization effect. Test, put in the box chemical indicator card (Shandong Xinhua Medical Instrument Factory) to reflect the sterilization effect. Test containers, points, in the next three layers, each layer in the 4 corner and the central press plum-like 5 points, each test a total of 15 points. Sterilization temperature of 121 ℃, the role of 20 minutes. Aluminum lunchbox test 32 times, a total of 30 box times, 30 instructions card has 12 from dark yellow, pass rate of 40%. Automatic opening and closing disinfection box tested 10 times, a total of 150 boxes,