我科的日产5151心电图机在使用近6年后突然热笔不热,经查是源于电源电路中的C_(1213)晶体管损坏,因购不到原管只得用同类元件代换,但一试机热笔阻丝即烧断。于是我们采用15W内热式电烙铁芯内瓷管(直径2~3mm)作笔杆,取300W电炉丝上相当于原热笔阻值(如10Ω)的1.1倍的一段,绕套均匀分布在瓷管上作发热体,一端与瓷管露出的引线相扎接,邻扎接处的阻丝弯制成热笔尖;然后设法从已坏的热笔体上敲掉笔杆,并把杆座改制成能接纳瓷管的凹状;
Our department’s Nissan 5151 ECG machine in the past 6 years suddenly hot pen is not hot, the investigation is derived from the power circuit C_ (1213) transistor damage, because the original tube can not buy the same type of components have to replace, but one Test machine hot wire resistance wire is blown off. So we use 15W inner electric iron core porcelain tube (diameter 2 ~ 3mm) for the pen, take 300W electric wire equivalent to the original hot pen resistance (such as 10Ω) of 1.1 times the section, around the sleeve evenly distributed in the porcelain On the heating element, one end of the porcelain tube exposed lead wire phase connection, the next junction of resistance wire bending made of hot pen tip; and then try to knock the pen from the bad hot pen body, Recessed shape of ceramic tube