论文部分内容阅读
我们参考了文献所述的光电头,结合我国实际,研制成WD-701型中型石英光电直读光谱仪,1971年在抚顺铝厂的非恒温室中运行后,许多同志对该仪器的温度补偿问题感到关切。为此,我们作了温度补偿实验和理论计算。实验和计算都证实了,在室温偏离调整出缝温度的±5℃范围内,温度补偿机构就能克服温度的影响。一、温度对中型石英光谱仪影响的计算
We refer to the literature described in the photoelectric head, combined with the actual situation in our country, developed WD-701 medium-sized quartz optical direct reading spectrometer, Fushun Aluminum Factory in 1971 non-temperature greenhouse operation, many comrades on the instrument temperature compensation Concerned. To this end, we made temperature compensation experiments and theoretical calculations. Experiments and calculations have confirmed that the temperature compensation mechanism can overcome the influence of temperature within the range of ± 5 ℃ when the room temperature deviates from the adjustment of the seam temperature. First, the temperature of the medium-sized quartz spectrometer impact calculation