故障 1 试剂探针漏液。在探针的运行路线上有明显的痕迹。出现漏液后 ,测定重复性明显不好。测定定值血清有时会出现明显偏离。检修 首先观察是那个探针漏液。仔细检查管道有无破损。如果管道完好 ,然后检查相应的注射器 ( PUMP)及电磁阀。一般仪器都配有备用电磁阀 ,如果怀
Fault 1 The reagent probe leaks. There are clear traces of the probe’s operating route. After the leakage occurred, the reproducibility was obviously not good. Determination of serum sometimes deviated significantly. Overhaul The first observation was that the probe leaked. Carefully inspect the pipe for damage. If the tubing is intact, then check the corresponding syringe (PUMP) and solenoid valve. General instruments are equipped with spare solenoid valve if you