A_(52),A_(53),A_(54)型检流计结构比较复杂,因此修理工艺要求就高一些,根据我多年的实际工作经验,在此提供几点修理经验。一、常见故障原因及修理方法 1、常见故障原因 (1)、通电时检流计不偏转可能①张丝或导流丝有损坏;②动圈和测量线路开路。 (2)、检流计偏转幅宽不对称①检流计平衡没有调整好;②有一根导流丝损坏;③导流丝焊接不平整或两根导流丝焊接后长短不一样;④水平位置不对。 (3)、检流计不回零位①由于严重冲击或张丝疲劳变质引起不回零。②可动线圈上有灰尘或其它杂质。 (4)、检流计灵敏度下降①大多数是由于接触不良、动圈短路;②磁场强度减弱等。
A 52, A 53 and A 54 type galvanometer structure is more complex, so the repair process requirements higher, according to my many years of practical work experience, here to provide some repair experience. First, the common causes of failure and repair methods 1, common causes of failure (1), Galvanic galvanometer can not be deflected ① wire or wire damage; ② moving coil and measuring circuit open. (2), the galvanometer deflection asymmetrical width ① galvanometer balance did not adjust well; ② a damaged wire; ③ wire welding uneven or two wire length after welding is not the same; ④ level Wrong position (3), Galvanometer does not return to zero ① due to severe impact or fatigue caused by Filament does not return to zero. ② movable coil dust or other impurities. (4), galvanometer sensitivity decreased ① Mostly due to poor contact, moving coil short circuit; ② magnetic field strength and so on.