故障现象 :水位正常 ,电加热管不工作。故障分析与检修 :该消毒器采用电加热管对水进行加热产生高压蒸气对医疗用品进行消毒。电热管安装在电锅炉内 ,电锅炉内水位由低水位控制器监测控制 ,保证锅炉水位低于一定值时能自动切断电加热管电源 ,停止加热。消毒器低水位控制器电路
Symptoms: The water level is normal and the electric heating tube does not work. Failure analysis and maintenance: The sterilizer uses electric heating tubes to heat the water to produce high-pressure steam to sterilize medical supplies. The electric heating pipe is installed in the electric boiler. The water level in the electric boiler is monitored and controlled by the low water level controller. When the boiler water level is lower than a certain value, the power of the electric heating pipe is automatically cut off and the heating is stopped. Sterilizer low water level controller circuit