SYSMEXF-8O0型血液分析仪是一种临床检验半自动血液学分析仪。由于其白细胞转换器的孔径只有IOOum,极易发生堵孔,而又不易清洗,直接影响白细胞计数结果。在临床工作中,笔者利用胰蛋白酶溶解孔周围的蛋白,效果很好。具体操作方法和结果:将堵塞的转换器拆下浸泡于l%的胰
The SYSMEXF-8O0 blood analyzer is a clinically-tested semi-automatic hematology analyzer. Because the pore size of its white blood cell converter is only IOOum, it is very easy to block pores, and it is difficult to clean, directly affecting the white blood cell count results. In clinical work, the author used trypsin to dissolve the protein around the pores, and the effect was very good. Specific methods and results: Remove the clogged converter and soak it in the 1% pancreas