我院一例20岁女性病人,因左肺下叶、舌叶支气管扩张;在气管内静脉复合全麻下行左肺下叶及舌叶切除,术中用心电图监护,切皮后按常规用KW65-3型电子管式高频电力止血。开始元火花往调整后止血效果好。术毕发现右大腿上外部及右臀部各有3×5cm和2×2cm电灼伤。原因分析KW65-3型电子管式高频电力备二个输出插口,一个“负”电极接地插口。正确操作是带输出线的主电极接输出插口,带引线的辅电
A 20-year-old female patient in our hospital suffered from bronchodilation of the left lower lobe and lingual lobe; left lower lobe and tongue resection under general anesthesia of endotracheal vein, electrocardiogram monitoring during operation, and conventional KW65- Type 3 electron tube high-frequency electric hemostasis. Start the yuan spark to adjust the hemostasis effect is good. After surgery, it was found that the external thigh and the right hip had 3 x 5 cm and 2 x 2 cm electric burns. Cause Analysis KW65-3 type electronic tube type high-frequency power is provided with two output sockets and one “negative” electrode grounding socket. The correct operation is to connect the main electrode with the output cable to the output jack, auxiliary power with lead wire