为了解鞍山地区医用诊断X射线摄影设备质量情况,2012年5-7月,我们对33家医疗单位的41台医用诊断X射线摄影设备进行了质量控制检测,结果报告如下。1对象与方法1.1对象鞍山地区33家医疗单位的41台医用诊断X射线摄影设备。1.2仪器和方法4000M+型X射线计量仪(经中国计量科学研究院校正),半值层测试铝板,垂直测试筒,准直测试板。按照《医用X射线诊断放射卫生防护及影像质量保证管理规定》
In order to understand the quality of medical diagnostic X-ray equipment in Anshan, from May to July 2012, we conducted quality control tests on 41 medical diagnostic X-ray equipment in 33 medical units. The results are reported as follows. 1 objects and methods 1.1 object Anshan 33 medical units in 41 medical diagnostic X-ray equipment. 1.2 instruments and methods 4000M + X-ray measuring instrument (calibrated by the China Institute of Metrology), half-value test aluminum, vertical test tube, alignment test board. In accordance with “Medical X-ray diagnosis of radiation health protection and image quality assurance management regulations”