故障现象:红细胞计数部分报警,仪器提示堵孔(Clog),T1=6.00S,T2=0.00S,所有红细胞、血小板的计数值均为空白,仪器无法使用。分析与检修:引起红细胞计数,堵孔报警(Clg,T1=6.005,T2=0.005)的原因主要来自以下三个方面:(1)真空泵压力小,引起T1延长。(2
Symptoms: The red blood cell count alarmed, and the instrument prompts a clog, T1=6.00S, T2=0.00S. All red blood cells and platelet counts are blank and the instrument cannot be used. Analysis and maintenance: Cause red blood cell count, blocking hole alarm (Clg, T1 = 6.005, T2 = 0.005) The main reasons come from the following three aspects: (1) vacuum pump pressure is small, causing T1 extension. (2