核探测器在高计数率情况下使用时,需要进行死时间修正,在精确测量中,高计数率还对谱分辨、峰面积、峰位置等产生影响,在实验中,用活度为微居级的60Co源作为待测源,用活度为毫居级的22Na源提供高计数率,通过移动22Na源与高纯锗探测器之间位置提供不同计数率的影响,用死时间表征计数率的大小,图1、2是测量的不同死时间下的两个脉冲幅度谱。图1、2都是选取
Nuclear detectors used in the case of high count rates, the need for dead time correction, in the precise measurement, the high count rate also spectral resolution, peak area, peak position and other effects in the experiment, with activity as a micro-residence Grade 60Co source as the source to be measured, a high count rate is provided by a 22Na source of activity level milli-scale, the count rate is varied by shifting the position between the 22Na source and the high-purity germanium detector, and the count rate is represented by the dead time Figure 1 and 2 show the two pulse amplitude spectra at different dead times. Figure 1, 2 are selected