Grolman(1972)介绍了非接触式眼压计(Noncontact Tonometer,NCT)。它以一般冲击角膜的气流,代替了仪器与眼组织的机械性接触。NCT包括三个系统即(1)对准系统;(2)光电压平监测系统和(3)气流系统。对准系统是一个供病人注视的内在目标,并有镜片组能使高度屈光不正的患者得到矫正而获得清楚的目标成像。视力差而不能固视时,可用对侧眼注视一个外在光源。检查者观察反映在病人角膜上的目标、并调节仪器位置使其重叠于固定的环上,只有当角膜的轴位、垂直位和侧位都校正后,才可进行眼压测量。光电压平监测系统包括一个能射出一束平行光线到角膜顶端的发送器,一个接受器和监测
Grolman (1972) introduced the Noncontact Tonometer (NCT). It is the general impact of the corneal airflow, instead of the mechanical contact with the eye tissue. The NCT consists of three systems, namely (1) Alignment System; (2) Photo Voltage Level Monitoring System and (3) Airflow System. The alignment system is an intrinsic target for patient gaze and has lens groups that allow patients with a high degree of refractive errors to be corrected for clear target imaging. Poor vision can not be fixed, the eye can be used to watch an external source of light. The examiner observes the target reflected in the cornea of the patient and adjusts the position of the instrument so that it overlaps the fixed ring. The measurement of intraocular pressure is performed only after the corneal axial, vertical and lateral positions are corrected. The optical voltage level monitoring system consists of a transmitter, a receiver and a monitor that emit a parallel beam of light to the corneal top