论文部分内容阅读
1故障现象在正常开机后,屏幕出现报警显示:VACUUMoutofRangeDuringthelastcyclePressENTERtocontinue意思是该次循环,真空值超出工作范围。按Enter键继续,执行后,重复上述报警。查看真空值,仪器显示:Vac-uumReading“0.3”mmHg,正常情况下应为“”。离池充满废液?
1 fault phenomenon After the normal boot, the screen shows an alarm: VACUUMoutofRangeDuringthelastcyclePress ENTERtocontinue means that the cycle, the vacuum value exceeds the working range. Press Enter to continue. After the execution, repeat the above alarm. Check the vacuum value, the instrument shows: Vac-uumReading “0.3” mmHg, under normal circumstances should be “”. Is the pool full of waste?