我院一台上海医用分析仪器厂生产的GD811型连续式比色计,由于具有连续吸液、排液、连续进样等优点而受检验人员的欢迎。但使用三年多后,发生一例故障:打开电源开关,仪器正常
论文部分内容阅读
我院一台上海医用分析仪器厂生产的GD811型连续式比色计,由于具有连续吸液、排液、连续进样等优点而受检验人员的欢迎。但使用三年多后,发生一例故障:打开电源开关,仪器正常,按下“进样开关”电源保险丝烧毁。按下“进样开关”电源保险丝烧毁,说明“进样开关”以后部分有严重漏电或短路现象。
GD811 continuous colorimeter produced by Shanghai Medical Analytical Instrument Factory in our hospital is welcomed by inspectors due to its advantages of continuous aspirating, draining, continuous feeding and so on. However, after using for more than three years, an example of failure occurred: turn on the power switch, the instrument is normal, press the “injection switch” power fuse burned. Press the “injection switch” power fuse burned, indicating “injection switch” part of a serious leakage or short circuit phenomenon.