故障现象:1.框架不能倾斜。2.通电后框架自动后倾-20℃。3.倾斜角数字显示与实际不符。故障分析和修复X-CT框架倾斜系统由控制按纽、综合方向/电流继电器、驱动机械和倾斜角度显示等几个部分组成,整个扫描允许倾斜范围为-20℃(后倾)到十20℃(前倾)。由图1知,框架倾斜角度的信息来自1K电位器,1458起一个求和放大器的作用,2脚处既有来自倾斜电位器滑动臂的电位,代表着扫描机架倾斜的角度;又有来自偏置电压+10V,其目的是使1458输出电压维持在A/D转换器的工作电压范围内。代表扫描框架倾斜角度的模拟电压经A/D转换后以数字形式显示在控制台荧光屏上。
Symptoms: 1. The frame can not be tilted. 2. After power automatically tilting the frame -20 ℃. 3. Tilting angle figures do not match with the actual. Failure analysis and repair X-CT frame tilt system consists of control buttons, integrated direction / current relays, driving machinery and tilt angle display and other components, the entire scan allows the tilt range of -20 ℃ (backward) to 10 ℃ 20 (Forward). From Figure 1, the frame tilt angle information from the 1K potentiometer, 1458 from the role of a summing amplifier, 2 feet both from the potentiometer sliding arm potential, representing the tilt angle of the scanning frame; also from The bias voltage is + 10V, the purpose of which is to maintain the 1458 output voltage within the operating voltage range of the A / D converter. The analog voltage representing the tilt angle of the scan frame is digitally displayed on the console screen after A / D conversion.