故障一:定标I不通过,仪器显示“吸定标I有气泡”。原因分析及排除:(1)看试剂容量是否低于2%,如低于2%应更换新的离子试剂包则可排除。(2)检查钾、钠、氯、混匀器、TCO2、参考电极组成的样品测量管道之间的“O”型密封圈是否压紧和位置是否准确。安装电极时,应把各电极的密封圈用卫生纸将水擦干净并且检查放置的位置即可排除。(3)吸样针与电极管道之间的联接管道是否松动。这个经常不为人们重视,如其他原因都排
Fault 1: calibration I do not pass, the instrument shows “suction standard I bubble”. Analysis and removal of reasons: (1) to see whether the reagent capacity is less than 2%, such as less than 2% should be replaced with a new ion reagent package can be ruled out. (2) Check potassium, sodium, chlorine, mixers, TCO2, reference electrode composed of sample measurement pipe between the “O” ring is pressed and the location is accurate. When installing the electrode, the electrode seal should be wiped clean with toilet paper and check the placement can be ruled out. (3) Whether the connecting pipe between the needle and the electrode pipe is loose. This often does not pay attention to people, such as other reasons are ranked