Electropolymerized o-phenylenediamine film is used as a functional coating for the immobilization of anti-S. aureus antibody on the surface of a gold-plated piezoelectric crystal, and this piezoelectric immunosensor is applied to detect S. aureus. The frequency shift (F = F20s - F380s, Hz) between the frequency at the 20th second (after the addition of sample, F20s) and that at 380 seconds later (F380s) was introduced to construct a calibration graph, and shortening of assay time was achieved. The S. aureus concentrations in the range of 105-109 cells/mL can be detected by this system.
Aureus antibody on the surface of a gold-plated Piezoelectric crystal, and this piezoelectric immunosensor is applied to detect S. aureus. The frequency shift (F = F20s - F380s, Hz) between the frequency at the 20th second (after the addition of sample, F20s) and that at 380 seconds later (F380s) was introduced to construct a calibration graph, and shortening of assay time was achieved. aureus concentrations in the range of 105-109 cells / mL can be detected by this system.