【摘 要】
:
该机为北京产200mA单床单管X线机,可以作透视、普通摄影、胃肠点片、断层等多项检查,最适用于县级及乡镇卫生院,但由于该机型为早期产品,其故障率相对较高,而较为常见的是控制电路中的限
论文部分内容阅读
该机为北京产200mA单床单管X线机,可以作透视、普通摄影、胃肠点片、断层等多项检查,最适用于县级及乡镇卫生院,但由于该机型为早期产品,其故障率相对较高,而较为常见的是控制电路中的限时其部分。故障现象:透视时有X线发生,做普通摄影及胃肠点片时无X线发生。?
This machine is a Beijing 200mA single-sheet X-ray tube machine, which can be used for perspective, ordinary photography, gastro-intestinal patches, faults, and many other inspections. It is most suitable for county-level and township hospitals, but because the model is an early product, The failure rate is relatively high, and the more common is the time-limited part of the control circuit. Symptoms: X-rays occur during fluoroscopy, and X-rays do not occur when normal photography and gastrointestinal spotting are performed. ?