上海医疗器械四厂生产的403型基础代谢测定仪采用了气相色谱的热导池检定器(TCD)的差示测量法,机体在呼吸空气时测定基础代谢率(BMR)和呼吸商(RQ)的数值。但是在较高海拔地区较低气压时使用该型仪器,得出的BMR值,必须加以校正,否则误差较大。为求得校正系数,制作了《t·p—μ表尺》,见附图。在昆明地区(海拔1966.4M)经500余例临床检测证实,本表尺较为精确,查读简便。现将其制作和使用方法简介如下:
The 403 Metabolometer, manufactured by Shanghai Medical Device Factory No. 4, uses a differential gas chromatograph (TCD) differential measurement method to measure the basal metabolic rate (BMR) and respiratory quotient (RQ) of the body when breathing air. The value of. However, when using this type of instrument at lower atmospheric pressure at higher altitudes, the resulting BMR value must be corrected or the error will be greater. In order to obtain the correction factor, a “t · p-μ gauge” was made, as shown in the attached drawings. In Kunming area (1966.4M above sea level) by more than 500 cases of clinical tests confirmed that the meter ruler is more accurate, easy to read. Now its production and use of methods are as follows: