目的分析重力因素对二维探测器阵列验证静态调强计划的影响,判断机架角度归为0°的测量方法是否安全可靠。方法在0°机架角和实际治疗机架角分别测量静态调强计划的剂量分布,以3mm范围内偏差<3%(3%3mm)标准进行γ分析,获得相对于参考剂量分布的通过率,分析通过率变化规律。分析两种方法测量的等中心点绝对剂量的差异。结果通过率的变化呈随机分布,96.9%的照射野偏差<2.5%。所有计划的85.7%绝对剂量偏差<2%,最大偏差为4.75%。结论使用二维探测器阵列在0°角进行调强计划的日常验证是安全可靠的。
Objective To analyze the influence of gravity on the validation of two-dimensional detector array to verify the static power leveling project, and to judge whether the measurement method of 0 ° angle is safe and reliable. Methods The dose distribution of the static intensity-modulated program was measured at 0 ° gantry angle and the actual treatment gantry angle respectively. The γ analysis was performed with a deviation of <3% (3% 3 mm) in the 3 mm range to obtain the passage ratio relative to the reference dose distribution , Analyze the change rule of passing rate. The differences between the absolute dose at the isocenter measured by the two methods were analyzed. The results of the pass rate changes were randomly distributed, 96.9% of the radiation field deviation <2.5%. 85.7% of all planned absolute dose deviations <2% with a maximum deviation of 4.75%. Conclusion The routine verification of an intensity modulation scheme using a two-dimensional detector array at an angle of 0 ° is safe and reliable.