VarianCL2 30 0C D直线加速器在停机后重新启动时有时会发生下列故障 :不能进入主菜单并且显示Data13或Da ta31故障信息。1 产生Data13的原因数据 13故障信息是“读取版本号识别时通讯失败” ,也就是控制计算机在初始化过程中不能与通讯处理器联系上时系统将直接进入C :VARIA
When the VarianCL2 30 0C D Linear Accelerator restarts after a stop, the following faults can sometimes occur: The main menu cannot be entered and the Data13 or Data31 fault message is displayed. 1 The reason for generating Data13 data 13 The fault message is “Communication failed when reading the version number identification.” That is, the control computer will not be able to contact the communication processor during the initialization process. The system will directly enter C:VARIA.