机型:瑞典 gambro、AK—10系列故障现象,机器透析正常。按下“热水消毒”按钮时,“热水消毒”指示灯不亮,不能进入热水消毒程序。同样,也不能进行药液消毒和排空。故障分析:根据故障现象和以往检修经验判断,其故障可能是:(1)“SAGS”(Safety Guard swith)压力变换器接管口处漏气。(2)“SAGS”压力变换器开关损坏。(3)三号板(一号微处理器)有故障。以上三种故障、尤其第二种情况较为常见。检修:(1)用大钳子将压力变换器“SAGS”接管口处的螺丝旋紧。然后,开机观察。故障如故,由此
Model: Sweden gambro, AK-10 series of symptoms, the machine dialysis normal. Press the “hot water disinfection” button, “hot water disinfection” light is not bright, can not enter the hot water disinfection program. Similarly, disinfectant and emptying can not be done. Failure Analysis: According to the fault phenomenon and previous maintenance experience to judge, the fault may be: (1) “SAGS” (Safety Guard swith) pressure transducer leakage at the nozzle. (2) SAGS pressure switch is damaged. (3) No. three board (No. 1 microprocessor) is faulty. The above three kinds of faults, especially the second case is more common. Overhaul: (1) Use a pair of pliers to tighten the screw on the pressure transducer “SAGS”. Then, start the observation. Trouble as before, thus