【摘 要】
:
故障现象开机后,监视器屏幕无图像。故障检修首先用信号发生器从监视器视频输入端子上输入一个信号,可观测到监视器屏幕上有图像出现,说明监视器工作正常,故障出现在主机电路上。
【机 构】
:
国防科技大学医院!长沙市,410073
论文部分内容阅读
故障现象开机后,监视器屏幕无图像。故障检修首先用信号发生器从监视器视频输入端子上输入一个信号,可观测到监视器屏幕上有图像出现,说明监视器工作正常,故障出现在主机电路上。用万用表测量电源部分基准电源电压值,发现-5V电源对地电压为0V,对地电阻为0Ω,说明电路中
Symptoms After the camera is turned on, there is no image on the monitor screen. Troubleshooting First, use the signal generator to input a signal from the video input terminal of the monitor. Observe that an image appears on the monitor screen, indicating that the monitor is working properly and that the fault appears on the host circuit. Using a multimeter to measure the voltage reference value of the power supply section, it was found that the voltage of the -5V power supply is 0V to ground, and the resistance to ground is 0Ω.
其他文献
γδT细胞与抗体一样能同抗原直接反应,也能同蛋白以外的物质进行反应。主要分布于小肠上皮、皮肤等处,与普通的T细胞功能不同。T细胞的抗原受体(也叫T细胞受体)有αβ型和γ
日本产CA300血球计数仪以其体积小、操作简单、故障率低深受基层医疗卫生单位的欢迎,现将工作中遇到的一特殊故障介绍给大家。故障现象;按装和调试CA300血球计数仪做空白时,红细胞(RBC)的显示值
故障现象:开机时按机柜右侧的reset键,reset灯亮,但在控制台上不能开机。故障分析与检修:检查低电压电源箱,发现继电器K2、K11、K15、K13、K16等均吸合,而K3不吸合,正常情况下K。也应吸合。检查K。继电器未发现损
故障现象一:拍出的多幅相片第1、2幅空白无图像,第3幅上迭加有多次曝光的图像,从第4幅至第15幅图像均正常,见图1。分析检修:打开照相机侧面板检查发现当拍完最后一幅图像,镜头从第15幅位置由
HG-3型火焰光度计是用于测量人的体液中钾、钢电解液浓度的仪器,它具有结构紧凑,操作简单,灵敏度高,读数稳定,试样量少等优点。其基本结构是由燃烧和气路系统、光学系统、信号处理
1997年 3~ 7月 ,在大鼠 D-半乳糖加速衰老模型基础上 ,注射鹅膏蕈氨酸 ( ibotenic acid,IBO)损毁脑 Meynert核制造了新的痴呆大鼠模型 ,并对其学习记忆能力的改变进行了初步观
节理的三维形貌特征是影响节理剪切力学行为的重要因素。为了深入研究三维形貌特征对岩石节理峰值抗剪强度的影响,制备花岗岩和红砂岩人工劈裂岩石节理试样,并在常法向应力条
同志们:在全国文化战线认真贯彻党的十四届六中全会精神,进一步深化文化体制改革.推进文化艺术事业发展的大好形势下,全国“文化长廊之歌”征歌颁奖会.今天在北京隆重召开了
本文介绍HP78系列床旁监护仪电源板中VR1稳压二极管变质失效后,对主机不工作故障进行原理上的分析及排除故障的过程。
This article describes the HP78 series bedside monitor pow
故障现象,百特550肾机开机后一直低温报警。分析检修:百特550肾机恒温控制系统由温度测量系统、温度控制板PC—4C,加热系统及高/低温报警系统等组成。为了缩小故障范围,我们先将机器开至冲洗