目的了解压力蒸汽生物监测快速阅读器未显示结果的原因,探讨正确使用方法。方法通过1例快速生物监测阅读器未显示结果的报告,系统分析原因。结果未显示结果的原因主要有生物指示剂未插入到位,未与底部荧光探测仪监视孔充分接触;阅读孔受到灰尘、药剂等的污染或阅读灯出现故障。结论应掌握快速生物阅读器的保养及操作方法,确保得出准确的监测结果。
Objective To understand the reasons why the pressure steam biomonitor rapid reader does not display the result, and to explore the proper use of the method. Methods A rapid biomonitor reader did not show the results of the report, the system analysis of the reasons. The results did not show the results of the main biological indicators are not inserted in place, not with the bottom of the fluorescence detector monitoring hole full access; reading hole by dust, chemicals, pollution or reading lamp failure. Conclusion The maintenance and operation of rapid biometric readers should be mastered to ensure accurate monitoring results.