日立7170A生化分析仪具有自动化程度高、精密度高、准确性好、重复性好等优点,此仪器一般故障中最常见的现象是“杯空白报警”。造成这一现象的原因众多,有反应杯正常污染,反应杯出现划痕或裂纹,有分离胶污染,供水滤网及反应槽排水过滤网的原因,还有电磁阀SV15堵脏,水质不达标,清洗液及光源灯的原因等等。经过7年的使用,我们对造成这一现象的众多原因及处理方法作一总结,供同行参考
Hitachi 7170A biochemical analyzer with a high degree of automation, high precision, good accuracy, good repeatability, the general failure of this instrument is the most common phenomenon “cup blank alarm.” This phenomenon is caused by many factors, such as the normal reaction of the cuvette, scratches or cracks in the cuvette, the contamination of the separation gel, the filter of the water supply filter and the discharge channel of the reaction tank, the blockage of the solenoid valve SV15 and the non-compliance of the water quality , Cleaning fluid and light source and so on. After seven years of use, we summarize many reasons for this phenomenon and how to deal with them for peer reference