德国蔡氏公司出的HFO-2型火花发生器是发射光谱分析用的高压火花光源,在国内尚有早期进口的这类仪器在使用。我们发现这种发生器在控制火花放电时会出现不能同步的现象,表观现象为放电时可以用耳朵听出有断续声,示波管上的放电波形不能湿示出稳定波形,时有时无,以致不能作分析用。这是由于同步马达中的同步开关不能断开、使继电器R3-RH94自始至终都通电,由Gr3整流出来的直流电压无法加到同步马达的转子线圈中去,马达始终作异步运行,同步断续器不能正常工作所致。发生这种故障时,可以拆下同步电机修理。但由于马达不易拆卸,不易调整,因此采取下述应急修理措施。考虑到同步马达是处于轻负荷起动状态,在起动阶段,转子中加入的直流工作电压的阻尼作用对起动性能影响不大,并且在起动期间的感应电动势还达不到烧毁整流元件的数值。鉴于上述理由,我们把同步
The HFO-2 spark generator from Chua’s company in Germany is a high-pressure spark light source for emission spectrum analysis, which is still being used in China. We found that this generator in the control of spark discharge can not be synchronized phenomenon, the apparent phenomenon of discharge can be heard with the ears intermittent sound, the oscilloscope discharge waveform can not be wet to show a stable waveform, sometimes when No, so can not be used for analysis. This is because the synchronous switch in the synchronous motor can not be disconnected, so that the relay R3-RH94 is energized from start to finish. The DC voltage rectified by Gr3 can not be added to the rotor coil of the synchronous motor, and the motor always performs asynchronous operation. Can not work properly. In the event of such failure, the synchronous motor can be removed for repair. However, due to the motor is not easy to disassemble, easy to adjust, so take the following emergency repair measures. Considering that the synchronous motor is in a light load starting state, the damping effect of the DC working voltage added in the rotor has little effect on the starting performance during the starting phase, and the induced electromotive force does not reach the value of the burned rectifier during starting. In view of the above reasons, we synchronize