论文部分内容阅读
本文介绍CGR—1477型碳玻璃(Carbon glass)温度计在1.5K—8K,0—2.83T磁场下进行温度修正的结果,并给出了温度修正的拟合公式,从而推出了温度修正与磁场的直接依赖关系。