CXL-101型色谱仪是国内最早生产的带管式炉裂解器的色谱仪。为了提高和完善仪器的性能,我们对裂解炉部分作了以下一些改进。 (1)改变汽化室、裂解炉和四通阀之间的相对位置。原仪器裂解炉位于仪器顶部中央,比主机高出10厘米,而四通阀在右下侧,比顶部低10厘米,其间管路较长,并完全暴露于空气中。这种结构不仅包装、运
CXL-101 chromatography is the earliest production of tube furnace cracker chromatograph. In order to improve and perfect the performance of the instrument, we made the following improvements to the cracking furnace. (1) to change the relative position of the vaporization chamber, pyrolysis furnace and four-way valve. The original instrument pyrolysis furnace is located at the center of the top of the instrument, 10 centimeters above the mainframe, while the four-way valve is 10 centimeters below the top on the lower right, with a long pipeline and complete exposure to the air. This structure is not only packed, shipped