论文部分内容阅读
故障现象装机正常使用半年余,在一次按正规操作规程使用过程中突然出现死机,操作面板触摸按键全部失灵,显示窗显示:ENTERVA-LUESORARMSIZE6OHVOL35.O,提示可以输人注射参数或准备注射,属正常显示。关机后再开机,故障依;日。在此之前曾出现过几次该现象,但关机后再开
Failure phenomenon Normal use of more than six months, a sudden crash occurred during the use of a regular operating procedure, the touch panel operating panel failure, display window display: ENTERVA-LUESORARMSIZE6OHVOL35. O, suggesting that you can input parameters or prepare injections, it is a normal display. Shutdown and then boot, the fault by; date. There have been several times before this phenomenon, but after the shutdown