论文部分内容阅读
利用 Na I O4鲁米诺流动注射化学发光法,以 8羟基喹啉掩蔽 M n2+ 以外的其它金属离子,柠檬酸钠掩蔽 M n2+ 和 Co2+ 以外的金属离子,达到同时测定 M n2+ 和 Co2+ 的目的。 M n2+ 离子的检测限达0194 ng/m l,r= 0.9984,线性范围(10- 9~10- 7)g/m l,回收率 88.0% ~107.6% ,测定 R S D 为 1.42% 。 Co2+ 离子的检测限达 1.717 ng/m l,r= 0.9993,线性范围(10- 8 ~10- 6)g/m l,加标回收率 87.8% ~1155% ,测定 R S D 为 1.96% 。
Flow-injection chemiluminescence with Na I O4-Luminol was used to mask other metal ions except M n2 + with 8-hydroxyquinoline. Sodium citrate masked metal ions other than M n2 + and Co2 + to achieve the simultaneous determination of M n2 + and Co2 + purpose. The detection limit of M n2 + was 0.194 ng / ml, r = 0.9984, and the linear range was from 10- 9 to 10-7 g / ml. The recoveries ranged from 88.0% to 107.6% R S D is 1.42%. The detection limit of Co2 + ion was 1.717 ng / m l, r = 0.9993, and the linear range (10 -8 ~ 10 -6) g / m l. The recoveries were 87.8% ~ 1155% R SD was determined to be 1.96%.