原锦州铁路局管内有医用诊断X线机46台经监测不合格的29台,占63%。改为锦州分局后有机器28台,其中不合格的12台,占43%。在多年监督监测工作中,发现一些值得重视的问题。现提出个人几点看法,供同志们参考。 1.有关主管与设计部门忽视“三同时”原则。X线机房设计审查多不邀请防护监督部门参加,而多由不懂X线防护的主管或基层单位提出设计方案。结果因设计不当,造成了难以挽回的被动局面。如某卫生所选址低洼,机房地面经常处于潮湿状态。雨水一多,机器下部泡在水里造成损坏。某医院机房未经防护监督部门审查擅自安在楼上,因
The original Jinzhou Railway Bureau tube medical diagnostic X ray machine 46 units failed to monitor 29 units, accounting for 63%. After the change to Jinzhou Branch 28 machines, of which 12 failed, accounting for 43%. During years of monitoring and monitoring work, some problems worth noting were discovered. Now put forward a few personal views for comrades reference. 1. Managers and design departments ignore the “three simultaneous” principle. X-ray machine room design review and more do not invite the protection and supervision departments to participate, but many do not understand the X-ray protection by the competent or grass-roots units proposed design. The result of improper design, resulting in a passive situation of irreplaceable. Low-lying sites such as a health clinic, the engine room floor often in wet conditions. More rain, the lower part of the machine soaked in water damage. A hospital room without supervision of the supervision department review without authorization upstairs, due